Page 2
LumaSense Technologies prohibits the duplication of any portion of this manual or the use thereof for any purpose other than the operation or maintenance of the equipment described in this manual, without the express written permission of LumaSense Technologies.
1.3 Limit of liability and warranty ................6 1.4 Unpacking the Instrument ..................6 1.5 Service Request, Repair, or Support ..............7 1.6 Shipments to LumaSense for Repair ..............7 1.7 Transport, Packing, Storage ................... 8 1.8 Disposal / decommissioning ................... 8 Introduction ........................
Page 4
6.3 Changing of optics ....................29 Data format UPP (Universal Pyrometer Protocol) ............31 Reference Numbers ....................... 35 8.1 Reference numbers instrument ................35 8.2 Reference numbers accessories ................35 Troubleshooting ......................37 Index ............................. 39 IS 12-TSP / IGA 12-TSP Manual Contents • iv...
Warning: The pyrometer is only to be used as described in this manual. It is recommended that you only use accessories provided by the manufacturer. In addition, signs and markings on the device is to be observed and maintained in legible condition. IS 12-TSP / IGA 12-TSP Manual General Information • 5...
LumaSense Technologies is not liable for any damages that arise from the use of any examples or processes mentioned in this manual or in case the content of this document should be incomplete or incorrect.
Email: support@lumasenseinc.com 1.6 Shipments to LumaSense for Repair All RMA shipments of LumaSense Technologies instruments are to be prepaid and insured by way of United Parcel Service (UPS) or preferred choice. For overseas customers, ship units air- freight, priority one.
Storing the instrument out of these conditions can cause damage or result in malfunction of the pyrometer. 1.8 Disposal / decommissioning Inoperable instruments must be disposed of in compliance with local regulations for electronic materials. IS 12-TSP / IGA 12-TSP Manual General Information • 8...
2 Introduction 2.1 Appropriate use IS 12-TSP and IGA 12-TSP are extremely precise and longer-term stable transfer standard pyrometers which can be used for the checking of calibration sources in the respective pyrometer spectral range. 2.2 Scope of delivery Instrument with selectable optics, PC measurement and evaluation software „InfraWin“, works certificate with 3 selected measuring points, user manual.
Page 10
Built-in parallax free thru-lens view finder; additionally, laser targeting light (max. power level < 1 mW, λ = 630 - 680 nm, CDRH class II) (at IS 12-TSP with temperature range 850 - 2520 °C only with view finder) Parameters: Adjustable at the instrument or via serial interface: ε...
Page 11
For temperature ranges of up to 5% of the measuring range an additional temperature drift of 0.2 °C/ °C (23 °C) should be added. The temperature drift is 0.05 °C/ °C (23 °C) for measured temperatures of 5% to 10% of the measuring range. IS 12-TSP / IGA 12-TSP Manual Introduction • 11...
Label with optics information Parameter indicator 10 Mounting holes for accessories Parallax free thru-lens view finder 12 Type label (bottom side of the Setting keys (operate with tipp of instrument) ball-point pen) Connection cable IS 12-TSP / IGA 12-TSP Manual Introduction • 12...
3.1 Electrical Installation The IS 12-TSP or IGA 12-TSP are powered by a voltage of either 24 V (optional 15 to 40 V DC or 12 to 30 V AC, 48 to 62 Hz). As the instrument is operating with heated and thermostatically controlled sensors, it needs a warm-up of up to 30 minutes to achieve the maximum accuracy.
When using a converter RS485 RS232 take care, that the converter is fast enough to receive the pyrometer’s answer to an instruction of the master. Most of the commonly used converters are too slow for fast measuring equipment. So it is recommended to use the LumaSense converter (order no. 3 852 430).
Never look directly into the laser beam. The beam and spot can be watched safely from side. • Make sure that the beam will not be reflected into eyes of persons by mirrors or shiny surfaces. IS 12-TSP / IGA 12-TSP Manual Controls and Installation • 15...
For reasons of different sizes and construction types of calibration sources, the installation brackets from the LumaSense accessories are suitable to only a limited extent for these wide-ranging requirements. You must therefore purchase or build a suitable alignment aid yourself.
3.5 Sighting For exact aiming to the object the pyrometers are equipped with a thru-lens view finder and an additional laser targeting light (exception: the IS 12-TSP with temperature range 850 to 2520 °C has only a view finder). 3.5.1...
(see 3.5, Sighting). For fixing the optics has to be turned clockwise. Note: The measuring object has to be bigger than or at least as big as the spot size of the pyrometer. IS 12-TSP / IGA 12-TSP Manual Controls and Installation • 18...
3.6.3 Other measuring distances For all other measuring distances the optics will be adjusted with help of the view finder or the laser targeting light (see 3.5, Sighting). IS 12-TSP / IGA 12-TSP Manual Controls and Installation • 19...
Page 20
To ensure consistent document formatting, this page was intentionally left blank. IS 12-TSP / IGA 12-TSP Manual Controls and Installation • 20...
LED’s indicate which pyrometer parameter is selected for reading or indicator: changing. Temperature display (°C / °F) Settings: °C The temperature can be displayed in °C or °F. °F IS 12-TSP / IGA 12-TSP Manual Settings / parameter descriptions • 22...
90% of the recorded temperature difference. In the “0.00” position (Software InfraWin: „min“), the device operates using its time constant. 10.00 s The dynamic exposition time adjustment prolongs the exposure time at the lower range limit. IS 12-TSP / IGA 12-TSP Manual Settings / parameter descriptions • 23...
The relay contact is connected directly to the pyrometer between pins J and K. This mode allows a new value to be established, after each impulse from the reset signal. IS 12-TSP / IGA 12-TSP Manual Settings / parameter descriptions • 24...
If the address of a pyrometer is unknown, it is possible to communicate with it using the global Address 99 (connect only one pyrometer). IS 12-TSP / IGA 12-TSP Manual Settings / parameter descriptions • 25...
2 °C min. or 1% of the span of the temperature sub range above the minimum temperature, the supply voltage (pin K) is connected to pin J. IS 12-TSP / IGA 12-TSP Manual Settings / parameter descriptions • 26...
The first time you load InfraWin 5, you will be prompted to select a default language. The InfraWin software is available in German, English, Spanish, French, Portuguese, and Chinese. Once installed, click Language/Languages if you would like to select another language. IS 12-TSP / IGA 12-TSP Manual Software InfraWin • 27...
Page 28
To ensure consistent document formatting, this page was intentionally left blank. IS 12-TSP / IGA 12-TSP Manual Software InfraWin • 28...
Focusable optics 1: color mark red / black / red Focusable optics 2: color mark red / blue / red Focusable optics 3: color mark red / green / red IS 12-TSP / IGA 12-TSP Manual Maintenance • 29...
Page 30
To ensure consistent document formatting, this page was intentionally left blank. IS 12-TSP / IGA 12-TSP Manual Maintenance • 30...
7 = external deletion storage: 2 = 0.05 s 5 = 5.00 s 8 = automatic deletion 3 = 0.25 s 6 = 25.00 s 9 = hold IS 12-TSP / IGA 12-TSP Manual Data format UPP (Universal Pyrometer Protocol) • 31...
Page 32
Ref. number: AAbn Output: XXXXXX (hex 6-digit) Serial number: AAsn Output: XXXX (hex 4-digit) Device type: AAna Output: „IS 12/94“, „IS 12/65“or „IGA 12/157“ (16 ASCII-characters) IS 12-TSP / IGA 12-TSP Manual Data format UPP (Universal Pyrometer Protocol) • 32...
Page 33
Parameters Device type / AAve Output: XXYYZZ (6-digit decimal) software version: XX = 07 (instruments ID for IS 12-TSP or IGA 12-TSP) YY = Month of software version ZZ = Year of software version Software version in AAvs tt.mm.yy XX.YY detail: tt = day;...
Page 34
To ensure consistent document formatting, this page was intentionally left blank. IS 12-TSP / IGA 12-TSP Manual Data format UPP (Universal Pyrometer Protocol) • 34...
Calibration source has a non- during alignment to the calibration uniform temperature source distribution Laser targeting light Laser targeting light fails Instruments temperature above 55 °C. Choose a farther position to the calibration source IS 12-TSP / IGA 12-TSP Manual Troubleshooting • 37...
Page 38
To ensure consistent document formatting, this page was intentionally left blank. IS 12-TSP / IGA 12-TSP Manual Troubleshooting • 38...
Technical Data 9 Installation, Electrical 13 Temperature display 22 Interface 25 Troubleshooting 37 Language 27 Unpacking the Instrument 6 Laser targeting light 15, 22 UPP data format 31 Legend 5 Liability 6 IS 12-TSP / IGA 12-TSP Manual Index • 39...
Page 40
Warm-up 13 Warranty 6 IS 12-TSP / IGA 12-TSP Manual Index • 40...
Need help?
Do you have a question about the IS 12-TSP and is the answer not in the manual?
Questions and answers