Download  Print this page

Performance Data According To En 45544-1 Part 1 And 2 - Honeywell GasAlertQuattro Operator's Manual

1, 2, 3, and 4 gas detector
Hide thumbs

Advertisement

GasAlertQuattro
Operator's Manual

Performance Data According to EN 45544-1 Part 1 and 2:

Time of response
Time of recovery
Alarm response time
Zero variation
Overall uncertainty
Lower limit of measuring range
Drift under zero gas (3 months)
Drift under standard test gas (3 months)
Maximum calibration period under test conditions
(Under operation conditions the calibration period may differ from value under test conditions)
34
Shop for Gas products online at:
Target Gas
13s
15s
4s
2 ppm (v/v)
8% of the measuring value
1 ppm (v/v)
1 ppm (v/v)
1 ppm (v/v)
3 months
CanarySense
www.
CO
10s
10s
4s
0,3 ppm (v/v)
2.8%
0,2 ppm (v/v)
0,2 ppm (v/v)
2,3 ppm (v/v)
3 months
.com
H
S
2
1.888.610.7664

Advertisement

loading

  Related Content for Honeywell GasAlertQuattro