Before reassembling the detector, ensure the sealing
surfaces on the front and rear shells is free of debris
and moisture.
Replace the front shell. Press the front and rear
shells together firmly to ensure a proper seal. Ensure
the front and rear shells have a tight, uniform 1/16 in.
(1.6 mm) seal on all sides of the detector.
Replace the six machine screws using 3-4 in. lbs
torque. Do not overtighten the screws. Replace the
battery pack.
CanarySense
www.

Specifications

Instrument dimensions: 13 x 8.1 x 4.7 cm
(5.1 x 3.2 x 1.5 in.)
Weight:
316 g (11.15 oz.) with rechargeable battery pack
338 g (11.92 oz.) with alkaline battery pack
Operating temperature: -20°C to +50°C (-4°F to +122°F)
Storage temperature: -40°C to +60°C (-40°F to +140°F)
Operating humidity: 10% to 100% relative humidity
(non-condensing)
Operating environment specifications for use according to
European performance approvals
(Oxygen and methane measurements)
Operating temperature range certified by BAM: -20°C to
+50°C
Operating humidity as tested by BAM: 5% r.H. to 95% r.H.
(Extended Range of Use for Temperature and Humidity as
Compared To EN 50104 (Oxygen performance) and
EN 67009-29-1 (LEL performance))
Storage temperature range tested by BAM: -25°C to +60°C
Operating pressure as tested by BAM: 80 kPA to 120 kPa
Storage duration: Two years from purchase date
Dust and moisture ingress: IP66/67 (with screw engaged)
Alarm setpoints: May vary by region and are user-defined