What You See
Control Test
result is out
of range (too
high or too
low).
Meter shows
dashes before
blood is applied.
Some display
segments do not
appear to be
working when
the meter
handle is all the
way out.
44
What It Means
• Control solution is past
expiration date or is past
the 6-month open use-
life date.
• 10-test disc is past
expiration date.
• Deteriorated test strip
due to heat or exposure
to moisture.
• Control solution not at
room temperature.
• Not enough control
solution drawn into test
strip.
Your meter may be dirty.
This may affect the way
you see your results.
Invalid test result.
What You Do
Run another control test.
If still out of range, retest
with a new 10-test disc and
control solution with an
acceptable expiration date
and open use-life date. If
still out of range, contact
the Bayer Diabetes
Helpline (see pg. 45).
Clean meter (see pg. 27). If
you continue to experience
difficulty, contact the Bayer
Diabetes Helpline (see
pg. 45).
If there is a difference in
your meter's display and
the one shown on page v,
contact the Bayer Diabetes
Helpline (see pg. 45).
Discard strip and retest.