Safety
Electromagnetic Emissions: Environment Guidance
Emissions Test
RF emissions, CISPR 11
RF emissions, CISPR 11
Harmonic emissions, IEC 61000-3-2
Voltage fluctuations/flicker
emissions, IEC 61000-3-3

Approved Cables for Electromagnetic Compliance

Cables connected to the system may affect its emissions. Use only the cable types and lengths
listed here.
WARNING
Using cables, transducers, and accessories other than those specified for use with the
system may result in increased emissions from, or decreased immunity of, the system.
Approved Cables
Cable
Transducer Cable Kit (C5-2 , L12-4)
56
Compliance
Group 1
Class A
Class A
Complies
Length
Philips Part
Number
1.25 m (4 ft)
453561806941
Electromagnetic Compatibility
Electromagnetic Environment
Guidance
The system uses RF energy only for its
internal function. Therefore, its RF
emissions are very low and are not
likely to cause any interference in
nearby electronic equipment.
The system is suitable for use in all
establishments, except domestic
establishments and those directly
connected to the public low-voltage
power supply network that supplies
buildings used for domestic purposes.