4 Configuration Settings Appendix
Configuring CO
MX800
Measurement Settings: Main Setup -> GM (or AGM) -> CO
MP5-90
CO
can be measured by an anesthetic gas analyzer (AGM, G1, or G5), measurement extension modules
2
only
(Microstream, Sidestream, or Mainstream), or the integrated CO
listed in this section only apply to CO
Factory Defaults
Item Name
etCO
High
2
etCO
Low
2
etCO
Alarms
2
imCO
High
2
imCO
Alarm
2
CO
2
Scale
Unit
Color
Humidity Corr.
Gas Analyzer CO
Humidity Corr. This setting determines the method used to correct the influence of water vapor in
the patient's breath on the CO
takes the partial pressure contributed by the water vapor into consideration and therefore results in lower
CO
readings when compared to DRY. For an exact definition and the formula used, see the section on
2
"Measurement Specifications" in the "Installation and Specifications" chapter of the Gas Analyzer
Instructions for Use.
Configuring awRR from Gas Analyzer
MX800
Measurement Settings: Main Setup -> GM (or AGM) -> awRR
MP5-90
AwRR can be derived from the CO
only
of the CO
measurement extension modules (Microstream, Sidestream, or Mainstream), or by the
2
integrated CO
derived from an anesthetic gas analyzer.
92
from Gas Analyzer
2
measured by an anesthetic gas analyzer.
2
Oper.
MX800,
Mode
MP5 - MP90 (H30)
C M Profile Adult
x
x 60 mmHg
x
x 25 mmHg
x
x On
x
x 4 mmHg
x
x Off
x
x et + im
x
x 50 mmHg
x
mmHg
x
White
x
Wet
Configuration Implications
2
reading. The options are Wet or Dry. Setting HumidtyCorr to Wet
2
measured by an anesthetic gas analyzer (AGM, G1, or G5), or by any
2
measurement in the MP5. The settings listed in this section only apply for the awRR
2
measurement in the MP5. The settings
2
Profile Pedi
(Airway Respiration Rate)
Measurement Settings
2
Profile Neo