Measurement Settings
Agent Id This setting is available for the M1026A/B AGM only. Setting Agent Id to Manual
requires that the user manually chooses the correct anesthetic agent during monitoring. If set to
Automatic, the gas analyzer automatically identifies the predominant anesthetic agent in the breathing
circuit. To configure the individual settings for each anesthetic gas for the M1026A/B AGM, you must
first set Agent ID to Manual. This allows you to select the individual gases. If you use Manual agent
identification, the agent that was selected last will become the default agent. If you want to use
Automatic agent identification, make sure to switch Agent ID back to Automatic when you are
done.
HAL (Halothan) Settings
Main Setup -> GM -> Setup Agent -> HAL
Main Setup -> AGM -> HAL
Factory Defaults
Item Name
inHAL High
inHAL Low
inHAL Alarms
etHAL High
etHAL Low
etHAL Alarms
Scale
Unit
Color
ISO (Isofluran) Settings
Measurement Settings:
Main Setup -> GM -> Setup Agent -> ISO
Main Setup -> AGM -> ISO
Factory Defaults
Item Name
inISO High
inISO Low
inISO Alarms
etISO High
etISO Low
Oper.
MX800,
Mode
MP5 - MP90 (H30)
C M Profile Adult
x
x 2.0%
x
x 0.0%
x
x On
x
x 1.6%
x
x 0.0%
x
x On
x
x 2.0
x
%
x
Red
Oper.
MX800,
Mode
MP5 - MP90 (H30)
C M Profile Adult
x
x 3.0%
x
x 0.0%
x
x On
x
x 2.5%
x
x 0.0%
4 Configuration Settings Appendix
Profile Pedi
Profile Pedi
Profile Neo
Profile Neo
89

Advertisement

Table of Contents

Comments to this Manuals

Symbols: 0
Latest comments: