4 Configuration Settings Appendix
VueLink Configuration Implications
Device Alarms The VueLink module itself generates INOPs, but does not generate alarms. If the
external device's alarms are on, the module transmits these to the monitor. Device Alarms lets you
select whether these alarms are indicated on the monitor (Accepted) or Ignored.
Note that the setting Device Alarms is not cloned between monitors. If you clone configurations
between IntelliVue monitors with VueLink plug-in modules, you must check that these settings are correct
and adjust them according to customer specifications, if needed.
Default Color The default color is the color used for any numerics that are not linked to a particular
wave, and for any waves for which no color is specifically configured. Note that this setting is not cloned
between monitors. If you clone configurations between IntelliVue monitors with VueLink plug-in
modules, you must check that these settings are correct and adjust them according to customer
specifications, if needed.
<Other Settings> All other VueLink settings are specific to the individual driver. They are stored
in the VueLink module and cannot be stored in the monitor configuration. See the documentation
supplied with the VueLink module for configuration information.
Configuring IntelliBridge
The default settings for the IntelliBridge (EC10) plug-in module and their configuration implications are
MX800
specific to individual IntelliBridge drivers. They are stored in the IntelliBridge EC10 module and cannot
MP40-90
be stored in the monitor configuration. See the documentation supplied with the IntelliBridge EC10
only
module and the related IntelliBridge Device Driver Instructions for Use.
Configuring the Gas Analyzer
General Gas Analyzer Settings
MX800
MP5-90
Measurement Settings: Main Setup -> GM (AGM)
option
All gas analyzer settings are available on monitors with option H30 only. The MP5 does not support the
H30 only
AGM (M1026A/B).
Factory Defaults
Item Name
1
Agent
CO
2
awRR
O
2
N
O
2
1
ISO
2
AGT
3
AGT1
3
AGT2
86
Oper.
MX800,
Mode
MP5 - MP90 (H30)
C M Profile Adult
x
x ISO
x
x not a setting, see "Configuring CO2 from Gas Analyzer" on page 92
x
x not a setting, see "Configuring awRR from Gas Analyzer (Airway Respiration Rate)" on page 92
x
x not a setting, see "O2 Settings" on page 87
x
x not a setting, see "N2O (Nitrous Oxide) Settings" on page 88
x
x not a setting, see "ISO (Isofluran) Settings" on page 89
x
x not a setting, see "AGT / AGT1 / AGT2 Settings" on page 88
x
x not a setting, see "AGT / AGT1 / AGT2 Settings" on page 88
x
x not a setting, see "AGT / AGT1 / AGT2 Settings" on page 88
Profile Pedi
Measurement Settings
Profile Neo

Advertisement