Measurement Settings
Configuring ΔTemp
Measurement Settings: Main Setup -> Measurements -> ΔTemp
ΔTemp is a derived measurement.
Item Name
First Temp
Second Temp
Δ
Temp
Measurement
Color
ΔTemp Configuration Implications
First Temp/Second Temp This setting is only available when Measurement is Enabled. It
lets you select two temperature labels to calculate the temperature difference. Note that none of Predictive
Temp labels can be selected for the ΔTemp calculation.
The formula used to calculate Δ Temp is: Δ Temp = First Temp - Second Temp.
Δ Temp / Measurement If Measurement is set to Enabled, the monitor automatically
switches the Δ Temp measurement On when both configured Temp sources are available. The user can still
switch Off Δ Temp in monitoring mode.
Set Measurement to Disabled, if you want the Δ Temp measurement to be permanently disabled,
which means that in monitoring mode it will not be possible to switch Δ Temp On.
Configuring VueLink
MX800
Measurement Setting: Main Setup -> Measurements -> VueLink X (or
MP40-90
Device Name)
only
Factory Defaults
Item Name
Device Alarms
Default Color
(Temperature Difference)
Oper.
MX800,
Mode
MP2 - MP90
X2
C M Profile
Profile
Adult
Pedi
x
x Trect
x
x Tblood
x
x Off
x
Enabled
x
Green
Oper.
MX800,
Mode
MP40 - MP90 (H10/20/40)
C M Profile Adult Profile Pedi
x
Accepted
x
Green
4 Configuration Settings Appendix
MP5T, MP5SC
MP5 (B10/B11/B14)
Profile
Profile
Profile
Neo
Adult
Pedi
Disabled
MX800,
MP40 - MP90 (H30)
Profile Neo Profile Adult Profile Pedi
Ignored
MP2/X2
Profile
Profile
Neo
Outdoor
Enabled
White
Profile Neo
85