Measurement Settings
EEG Montages Configuration Implications
Select Montage Lets you select the default montage the monitor uses when the EEG measurement
is started.
Configuring BIS
MX800
Measurement Settings: Main Setup -> Measurements -> BIS
MP20-90
only
Factory Defaults
Item Name
SQI
EMG
SR
1
Bursts
SEF
TP
Scale
Show Gridlines
Filters
Low Filter
High Filter
Notch Filter
High Limit
Low Limit
Alarms
Cont. Imp. Check
Smoothing Rate
Color
BIS
1.Numeric available with BISx module only.
BIS Configuration Implications
EMG This setting lets you switch the EMG numeric On or Off. The EMG (Electromyographic Activity)
numeric reflects the electrical power of muscle activity and high frequency artifacts.
SR lets you switch the SR numeric On or Off. The SR (Suppression Ratio) is the percentage of time
over the last 63-second period during which the EEG is considered to be in a suppressed state.
(Bispectral Index)
Oper.
MX800,
Mode
MP20 - MP90 (H10/20/40)
C M Profile Adult Profile Pedi
x
On
x
On
x
On
x
On
x
Off
x
Off
x
100uV
x
No
x
On
x
2 Hz
x
70 Hz
x
On
x
70
x
20
x
On
x
not applicable, this is not a setting, see "BIS Configuration Implications"
x
30 sec
x
Yellow
x
x not applicable, the BIS On/Off state is not a setting, see "BIS Configuration Implications".
4 Configuration Settings Appendix
MX800,
MP20 - MP90 (H30)
Profile Neo
Profile Adult Profile Pedi
Off
15 sec
Magenta
Profile Neo
81