4 Configuration Settings Appendix
RRspir Configuration Implications
Apnea Time The apnea alarm is a high priority red alarm used to detect apneas. The Apnea Time
defines the time period between the point where the monitor cannot detect any respiration activity and the
indication of the apnea alarm.
RRspir This setting lets you switch the RRspir measurement On or Off.
Configuring EEG
MX800
Measurement Settings: Main Setup -> Measurements -> EEG
MP40-90
only
Factory Defaults
Item Name
TP
SEF
MDF
PPF
Delta
Theta
Alpha
Beta
SEF Threshold
Numeric Average
Wave Scale
Show Gridlines
Low Filter
High Filter
Buffer A
Buffer B
Buffer C
Smoothing CSA
Impedance Limit
Color
EEG
EEG Configuration Implications
TP This setting lets you switch the TP numeric On or Off. The TP (Total Power) numeric indicates the
power in the measured frequency band.
SEF lets you switch the SEF numeric On or Off. The SEF (Spectral Edge Frequency) is the frequency
below which a configurable percentage (set by the SEF Threshold) of the Total Power is measured.
78
Oper.
MX800,
Mode
MP40 - MP90
C M Profile Adult
x
x On
x
x On
x
x Off
x
x Off
x
x Off
x
x Off
x
x Off
x
x Off
x
90%
x
8 sec
x
x 100uV
x
No
x
x 0.5 Hz
x
x 30 Hz
x
not applicable, this setting is stored in the Monitor Settings Block: see "Configuring CSA
Buffers" on page 166.
x
x
x
On
x
x 5 kOhm
x
Yellow
x
x not applicable, the EEG On/Off state is not a setting, see "EEG Configuration
Implications".
Profile Pedi
Measurement Settings
Profile Neo