Philips IntelliVue MX800 IntelliVue MP2 Configuration Manual Page 77

Patient monitor..
Hide thumbs

Advertisement

Measurement Settings
Spirometry Gas Compensation Settings
Measurement Settings: Main Setup -> Measurements ->
Spirometry -> Gas Compensation
Factory Defaults
Item Name
Mode
Balance Gas
Inspired O
2
Inspired Agent
Inspired Temp
Unit
Spirometry Configuration Implications
Mode Use this setting to choose the gas compensation mode. Select Manual to manually enter gas
concentrations or Gas Analyzer to derive gas concentrations from the Philips gas analyzer.
Note: Gas concentrations from the gas analyzer are only available for Philips gas analyzers, not for devices
connected via a Vuelink or IntelliBridge plug-in module. If gas concentrations from the gas analyzer are
selected but not all data is available, the missing data is taken from manually entered values. In case of
invalid data or no data at all, the INOP message SPIRO GAS COMPENS? is displayed.
Balance Gas This setting lets you select the type of balance gas used. Choices are N
Inspired O
match the concentration of inspired O
Configuring RRspir
MX800
Measurement Settings: Main Setup -> Measurements ->
MP40-90
Spirometry -> RRspir
only
Factory Defaults
Item Name
High Limit
Low Limit
Apnea Time
RRspir Alarms
RRspir
Oper.
MX800,
Mode
MP40 - MP90
C M Profile Adult
x
x Manual
x
x N
2
x
x 30%
x
x 0.0%
0
x
x 25
C
0
x
C
/ Inspired Agent / Inspired Temp These settings can be adjusted to
2
and anesthetic agent, as well as the temperature of the inspired gas.
2
(Respiration from Spirometry)
Oper.
MX800,
Mode
MP40 - MP90
C
M Profile Adult
x
x 30 rpm
x
x 8 rpm
x
x 20 sec
x
x On
x
x On
4 Configuration Settings Appendix
Profile Pedi
Profile Pedi
Profile Neo
, and N
0.
2
2
Profile Neo
60 rpm
30 rpm
77