Philips IntelliVue MX800 IntelliVue MP2 Configuration Manual Page 76

Patient monitor..
Hide thumbs

Advertisement

4 Configuration Settings Appendix
Spirometry AWP (Airway Pressure) Settings
Measurement Settings: Main Setup -> Measurements ->
Spirometry -> AWP
Factory Defaults
Item Name
PIP High
PIP Alarms
PEEP High
PEEP Low
PEEP Alarms
Scale
Color
Spirometry AWV (Airway Volume) Settings
Measurement Settings: Main Setup -> Measurements ->
Spirometry -> AWV
Factory Defaults
Item Name
MVexp High
MVexp Low
MVexp Alarms
MV
TV
Scale
Color
Spirometry Configuration Implications
MV This setting lets you choose the measured components for minute volume (inspiratory, expiratory,
inspiratory + expiratory or off). If set to Off, there will be no alarming for minute volume.
TV This setting lets you choose the measured components for tidal volume (inspiratory, expiratory,
inspiratory + expiratory or off). If set to Off, there will be no alarming for tidal volume.
76
Oper.
MX800,
Mode
MP40 - MP90
C M Profile Adult
x
x 40 cmH2O
x
x On
x
x 25 cmH2O
x
x 0 cmH2O
x
x On
x
x 40 cmH2O
x
White
Oper.
MX800,
Mode
MP40 - MP90
C M Profile Adult
x
x 8.0 l/min
x
x 4.0 l/min
x
x On
x
x exp + in
x
x exp + in
x
x 800 ml
x
White
Profile Pedi
25 cmH2O
40 cmH2O
Profile Pedi
4.0 l/min
2.5 l/min
200 ml
Measurement Settings
Profile Neo
20 cmH2O
20 cmH2O
Profile Neo
0.8 l/min
0.4 l/min
50 ml