Measurement Settings
Resp This setting lets you switch the Resp measurement On or Off. If Resp is switched Off, the small
current applied to the Resp electrodes to enable the impedance measurement is switched off.
Detection The respiration detection level can be configured to be set either automatically or
manually. For further information, see the section on "Changing Resp Detection Modes" in the monitor's
Instructions for Use.
Configuring Spirometry
MX800
Measurement Settings: Main Setup -> Measurements -> Spirometry
MP40-90
Factory Defaults
only
Item Name
No Al. til Breath
Color
Spirometry Configuration Implications
No Al. til Breath If set to set On, the monitor suppresses alarms from the Spirometry module
until it detects that a patient has been connected to the module (when breathing is detected).
Spirometry AWF (Airway Flow) Settings
Measurement Settings: Main Setup -> Measurements ->
Spirometry -> AWF
Factory Defaults
Item Name
Scale
Color
Oper.
MX800,
Mode
MP40 - MP90
C M Profile Adult
x
On
x
White
Oper.
MX800,
Mode
MP40 - MP90
C M Profile Adult
x
x 150 l/min
x
White
4 Configuration Settings Appendix
Profile Pedi
Profile Pedi
100 l/min
Profile Neo
Profile Neo
20 l/min
75