4 Configuration Settings Appendix
Factory Defaults
Item Name
High Limit
Low Limit
Apnea Time
Alarms
awRR
awRR Configuration Implications
Apnea Time The apnea alarm is a high priority red alarm used to detect apneas. The Apnea Time
defines the time period between the point where the monitor cannot detect any respiration activity and the
indication of the apnea alarm.
awRR This setting lets you switch the awRR measurement from CO
Configuring Resp
Measurement Settings: Main Setup -> Measurements -> Resp
Factory Defaults
Item Name
High Limit
Low Limit
Apnea Time
Alarms
Resp
Detection
Color
Resp Configuration Implications
Apnea Time The apnea alarm is a high priority red alarm used to detect apneas. The Apnea Time
defines the time period between the point where the monitor cannot detect any respiration activity and the
indication of the apnea alarm.
74
Oper.
MX800,
Mode
MP2 - MP90
X2
C M Profile Adult
x
x 30
x
x 8
x
x 20 sec
x
x On
x
x On
(Impedance Respiration)
Oper.
MX800,
Mode
MP5 - MP90 (H10/20/40)
MP2/X2
C M Profile
Profile
Adult
Pedi
x
x 30
x
x 8
x
x 20 sec
x
x On
x
x On
x
x Auto
x
Yellow
Profile Pedi
On or Off.
2
MX800,
MP5 - MP90 (H30)
Profile
Profile
Profile
Neo
Adult
Pedi
100
30
Off
White
Measurement Settings
Profile Neo
100
30
MP2/X2
Profile
Profile
Neo
Outdoor
30
8
On
White