Philips IntelliVue MX800 IntelliVue MP2 Configuration Manual Page 69

Patient monitor..
Hide thumbs

Advertisement

Measurement Settings
SO
Configuration Implications
2
Label> The On/Off state of the SO
<SO
2
automatically switched On when an SO
Light Intensity Set this to Off if you do not want the Light Intensity Indicator to be displayed
next to the SO
Catheter Factor Depending on the probe/catheter in use, you may need to enter a catheter
correction factor. The appropriate correction factor is indicated in the 'Accessories' chapter of the
IntelliVue Instructions for Use or in the catheter documentation. The Catheter Factor can only be
entered during an in-vivo calibration and is stored in the Optical Module. Note that this is not a
measurement setting that can be permanently stored in the monitor configuration.
Hb/Hct Entry This setting determines the lab value that should be used for the in-vivo calibration.
Available choices are Hct[%], Hb[g/dl], or Hb[mmol/l].
Hct(or Hb) Depending on the configuration of the setting Hb/HCT Entry, this lets you enter either
the Hct or the Hb value obtained from the laboratory analysis. Hct(or Hb) can only be entered during an
in-vivo calibration and is stored in the Optical Module. Note that this is not a measurement setting that
can be permanently stored in the monitor configuration.
Calibration Value This shows the calibration value resulting from the last in-vivo calibration. It is
stored in the Optical Module and is not a measurement setting that can be permanently stored in the
monitor configuration.
Configuring Sp-vO
MX800
Measurement Settings: Main Setup -> Measurements -> Sp-vO
MP40-90
Sp-vO
is a derived measurement. Oxygen extraction is the difference between the measured SpO
2
only
SvO
values. To calculate the Sp-vO
2
SpO
source which can be set to any of the available SpO
2
Factory Defaults
Item Name
SpO
Source
2
Sp - vO
2
Measurement
Color
Sp-vO
Configuration Implications
2
Source This setting is only available when Measurement is Enabled. The formula used to
SpO
2
calculate Oxygen Extraction is: Sp-vO
source used for the calculation. Possible sources are: SpO
2
numeric.
2
(Oxygen Extraction )
2
, the monitor needs an SO
2
Oper.
MX800,
Mode
MP60 - MP90
C M Profile Adult
x
x SpO
2
x
x Off
x
Enabled
x
Green
= SpO
2
4 Configuration Settings Appendix
measurement cannot be preconfigured. SO
2
transducer is connected to the monitor.
source (must be labeled SvO
2
labels except %SpO
2
Profile Pedi
Source - SvO
. SpO
2
2
, SpO
pr, SpO
2
2
2
2
T.
2
Profile Neo
Source determines the SpO
2
po, SpO
r, SpO
l.
2
2
2
is
and
2
) and an
2
2
69