4 Configuration Settings Appendix
SVR Configuration Implications
Label This setting is only available when Measurement is Enabled. It lets you select whether the
SVR or the SVRI (indexed SVR) is displayed. It is not possible to display both values at one time.
Arterial Source This setting is only available when Measurement is Enabled. Possible
sources are: ABPm, ARTm, AoM, UAPm, FAPm, BAPm. The formula used to calculate SVR is:
Set CVP This setting is only available when Measurement is Enabled. It defines a value to be used
in place of CVP to calculate the SVR if no measured CVP is available. It can be set between 0 and 16
mmHg.
SVR / Measurement If Measurement is set to Enabled, the monitor automatically switches
the SVR measurement On when CCO and the set arterial source is available. The user can still switch Off
SVR in monitoring mode.
Set Measurement to Disabled, if you want the SVR measurement to be permanently disabled,
which means that in monitoring mode it will not be possible to switch SVR On.
Note that the settings made here do not affect the SVR calculated in the Hemo Calcs window.
Configuring SO
The configuration settings for SO
SO
, SvO
2
MX800
Measurement Settings: Main Setup -> Measurements -> <SO
MP40-90
only
Factory Defaults
Item Name
High Limit
Low Limit
Alarms
SO
2
Light Intensity
Catheter Factor
Hb/Hct Entry
1
Hct [%]l
Calibration Value
Color
1. Setting available for the M1011A SO
68
SVR = 79,96
(Intravascular Oxygen Saturation)
2
can be set individually for each label - SO
2
, and ScvO
Settings
2
2
Oper.
M1011A: MX800, MP40 - MP90
Mode
M1021A: MX800, MP60 - MP90
C M Profile Adult
x
x 80%
x
x SvO
: 60%
2
SO
/ScvO
2
x
x On
x
x not applicable, the SO
Implications".
x
On
1
x
x not applicable, this is not a setting, see "SO2 Configuration Implications".
1
x
x Hct[%]
x
x not applicable, this is not a setting, see "SO2 Configuration Implications".
x
x not applicable, this is not a setting, see "SO2 Configuration Implications".
x
Yellow
module only.
2
(
Arterial Source
mean
--------------------------------------------------------------------------------------- -
CCO
Profile Pedi
: 70%
2
On/Off state is not a setting, see "SO2 Configuration
2
Measurement Settings
)
CVP
mean
, SvO
and ScvO
2
2,
Label>
2
Profile Neo
.
2