Philips IntelliVue MX800 IntelliVue MP2 Configuration Manual Page 58

Patient monitor..
Hide thumbs

Advertisement

4 Configuration Settings Appendix
1
P
, ABP, ART, Ao, UAP, FAP, BAP, P1, P2, P3, P4 Settings
Measurement Settings: Main Setup -> Measurements -> <Press Label>
Factory Defaults
Item Name
Alarms from
Sys. High
Sys. Low
Dia. High
Dia. Low
Mean High
Mean Low
Alarms
ABP (or other label)
Pulse (<Press Label>)
Scale
Mean Only
Filter
Mercury Cal
Artifact Suppr.
Unit
Color
Extreme Alarms
Δ Extreme High
Δ Extreme Low
Sys. High Clamp
Sys. Low Clamp
Dia. High Clamp
Dia. Low Clamp
Mean High Clamp
Mean Low Clamp
1.The label P will be interpreted as P1 by the IntelliVue Information Center. It is therefore not recommended to use P and P1
simultaneously on the IntelliVue monitor when connected to an IntelliVue Information Center.
58
Oper.
MX800,
Mode
MP5 - MP90 (H10/20/40)
MP2/X2
C M Profile
Profile
Adult
Pedi
x
x Systolic
x
x 160
120
x
x 90
70
x
x 90
70
x
x 50
40
x
x 110
90
x
x 70
50
x
x On
x
x not applicable, the Pressure On/Off state is not a setting, see "Invasive Pressure
Configuration Implications" on page 61.
x
x not applicable, see "Configuring Pulse from Invasive Pressure" on page 63.
x
x 150
100
x
x No
x
12 Hz
x
Yes
x
60 sec
x
mmHg
x
Red
x
Disabled
x
15
10
x
15
10
x
190
140
x
80
60
x
100
80
x
45
35
x
125
100
x
65
45
MX800,
MP5 - MP90 (H30)
Profile
Profile
Profile
Neo
Adult
Pedi
90
180
55
70
60
20
70
36
100
5
5
105
45
65
75
15
75
30
Measurement Settings
MP2/X2
Profile
Profile
Neo
Outdoor
160
90
90
50
110
70
150
White
15
15
190
80
100
45
125
65