4 Configuration Settings Appendix
Factory Defaults
Item Name
Oper.
Mode
C M Profile
Desat Delay
x
NBP Alarm Suppr.
x
Extd. Auto OnOff
x
Color
x
1.Profile SpotCheck for English language software only.
2.Setting is available for the CL Sp0
a Short Range Radio (SRR) connection.
3.Setting is only applicable for CL Sp0
4.Setting is only applicable for monitors with FAST SpO
5.Setting is only applicable if SmartAlarmDelay is set to On.
SpO
Configuration Implications
2
Label> The On/Off state of the SpO
<SpO
2
automatically switched On when an SpO
Mode This setting is only applicable for monitors that have a Short Range Radio (SRR) interface
installed. It is only available for the label %SpO
measurements from the telemetry device to be made on request and not continuously, helping to save the
telemetry device's battery power when it is connected to a monitor via a short-range radio link.
To ensure there is no gap in SpO
short range radio transmission, the SpO
situation unless Manual mode is set in both the telemetry device and the monitor.
Perfusion If Perfusion is switched Off, Perfusion is not measured and the Perf numeric
disappears from the Screen. Note that you will only see the Perfusion numeric on the Screen if
Perfusion is switched On, and a Perf numeric is configured on the Screen.
Signal Quality Set Signal Quality to Off, if you don't want the signal quality indicator to
be displayed next to the SpO
with FAST SpO
50
MX800, MP2 - MP90
MP5T, MP5SC
X2
20 sec
On
Disabled
Cyan
pod, and also for the label %SpO
2
pod if Mode is set to Auto.
2
(Option A01) and OxiMax compatible SpO
2
2
measurements when moving from standard telemetry transmission to
2
mode will be automatically switched to Continuous in this
2
numeric on the Screen. Note that this setting is only applicable for monitors
2
(Option A01) and OxiMax compatible SpO
2
T, and on monitors that are currently connected to a telemetry device via
2
measurement cannot be preconfigured. SpO
2
sensor is connected to the monitor.
T. Configure Mode to Manual to allow %SpO
2
(Option A02).
2
Measurement Settings
MP5SC,
MP2/X2
MP5#P05
Disabled
White
(Option A02).
2
is
2
T
2