Philips IntelliVue MX800 IntelliVue MP2 Configuration Manual Page 41

Patient monitor..
Hide thumbs

Advertisement

Measurement Settings
Configuring the System Pulse
Measurement Settings: Main Setup -> Measurements -> Pulse
Factory Defaults
Item Name
High Limit
Low Limit
Pulse Alarms
AlarmSrc (ECG/AR)
System Pulse
QRS Volume
Δ ExtrTachy
Tachy Clamp
Δ ExtrBrady
Brady Clamp
Alarms Off
AlarmSource Sel.
PulseAlarms Tele
System Pulse Configuration Implications
System Pulse The System Pulse setting allows you to configure the measurement source for the
System Pulse.
The pulse rate chosen as system pulse:
• is monitored as system pulse and generates alarms when you select Pulse as the active
Alarm Source
• is sent via the network to the Information Center, if available
• is trended in the HighRes Trends and stored in the monitor's databases.
The choices are SpO
Ao, PAP, UAP, FAP, BAP, and Auto. If you select Auto, the monitor automatically chooses a
pulse rate to be used as system pulse. It looks through the list from top to bottom and activates the first
pulse rate that is switched on and available.
Oper.
MX800,
Mode
MP5 - MP90 (H10/20/40)
MP5T, MP5SC
MP2/X2
C M Profile
Profile
Adult
Pedi
x
x not applicable, see "Configuring ECG/Pulse Alarms" on page 38.
x
x
x
x
x
x
x
x SpO
2
x
x not applicable, see "Configuring User Interface Settings" on page 126.
x
not applicable, see "Configuring ECG/Pulse Alarms" on page 38.
x
x
x
x
x
x
, SpO
pr, SpO
po, SpO
2
2
2
4 Configuration Settings Appendix
MX800,
MP5 - MP90 (H30)
Profile
Profile
Profile
Neo
Adult
Pedi
Auto
r, SpO
l, %SpO
2
2
MP2/X2
Profile
Profile
Neo
Outdoor
SpO2
T, P, ABP, ART,
2
41