4 Configuration Settings Appendix
Configuring ECG/Pulse Alarms
Measurement Settings:
Main Setup -> Measurements -> ECG -> AlarmSrc(ECG/AR), or
Main Setup -> Measurements -> Pulse -> AlarmSrc(ECG/AR)
Factory Defaults
Item Name
Alarms Source
ECG/Arrhy Alarms
Pulse Alarms
High Limit
Low Limit
Δ ExtrTachy
Tachy Clamp
Δ ExtrBrady
Brady Clamp
Alarms Off
AlarmSource Sel.
PulseAlarms Tele
Inop "PULSE NO AL" x
1.Profile SpotCheck for English language software only.
ECG/Pulse Alarms Configuration Implications
Alarms Source In most cases the heart rate and Pulse numerics are identical. In order to avoid
simultaneous alarms on heart rate and Pulse, the monitor uses either ECG or Pulse as its active alarm
source. The Alarm Source setting lets you choose ECG, Pulse or Auto as the source of heart-
related rate alarms.
38
Oper.
Mode
C M
C M
x x ECG/Arrhythm
Auto
x x On
x x Off
x x 120 160 200
x x 50
75
100
x
20
x
200 220 240
x
20
x
40
50
40
x
Enabled
x
Enabled
x
Enabled
Enabled
Profile
Auto
60
80
Measurement Settings
Auto
120
120
50
50
200
200
40
40