Measurement Settings
4 Configuration Settings Appendix
ECG AL. Off Inop If ECG alarms are off or Pulse is selected as active alarm source, the INOP
ECG/ARRH ALARM OFF is shown permanently. If you do not want this INOP to appear, you must set
ECG AL. Off Inop to Off. If you want the ECG/ARRH ALARM OFF INOP to be automatically
escalated to a yellow alarm after a fixed time, configure it to one of the available choices: Yellow @2h,
Yellow @4h, Yellow @6h, or Yellow @8h.
Fallback If Fallback is configured On and there is a LEAD OFF INOP in the primary lead (and in
the secondary lead, if you are using multi-lead monitoring) for longer than 10 seconds, and if another lead
is available, this available lead automatically becomes the primary lead. This is known as lead fallback.
When the Leads Off condition is corrected, the leads are automatically switched back.
37