7 Telemetry Device and CL Pods Use Models
• Alarms are independent at the bedside and Information Center.
• While ECG Source is the telemetry device the ECG Out on the Intellivue Patient Monitor is not
functional.
• The settings of the devices are not synchronized.
N O TE
Telemetry devices Rev. C.00.xx support TAAP and WTAAP. Telemetry devices Rev. D. support WTAAP
and Cableless Measurements.
Configuration: Use Model 3
Required Configuration on MP2, X2, MP5
Set Network setting TAAP to (for wired or wireless)
Assign Equipment label
SRR (for wireless) (in Setup Hardware)
SRR Channel (for wireless) (in Setup Hardware)
Recommended setting for Tele Discharge (in Global
Settings)
Tele Unassign
Required Hardware on MP2, X2, MP5
SRR interface or TAAP connector
IntelliVue Instrument telemetry (IIT) interface
Required Configuration at Information Center for MP2, X2, MP5
Equipment label requirement
Required Configuration at Information Center for Telemetry
Equipment allowed
Equipment label requirement
Required Hardware for Telemetry
Short Range Radio Adapter - SRRA or TAAP Cable
Troubleshooting FAQ for Use Model 3
There was a V Tach alarm at the monitor but not at the IIC, why did this happen?
• Alarms and settings are independent between the monitor and the telemetry device. Learning happens
independently, the beat templates that ST/AR stores at the IIC are not replicated to the monitor when
the assignment occurs.
I assigned the telemetry device with SRR to the monitor at the bedside and my bedside ECG showed up correctly,
then my patient went out of the room for a test. When my patient came back from the test the telemetry device
does not immediately change back to the bedside ECG, what can I do?
• The telemetry device searches for wireless radios for 30 seconds on startup or when the check button is
pressed to preserve battery life. You can speed up the discovery process upon return by pressing the
check button on the telemetry device when you get back to the patient's room.
The assignment did not stick while my patient went to X-ray, what should I do to make sure this does not happen
in the future?
298
Telemetry Device Use Models
Enabled
Not used in this configuration
On
11-26 (Service Personnel Only)
OnDevChange
1 min
yes
IIT cannot be used
none needed
IntelliVue Telemetry only
Telemetry Label - ITS