7 Telemetry Device and CL Pods Use Models
• When the devices are paired and the cable is disconnected or the short range radio link is lost, the screen
switches to the embedded telemetry device data window automatically once the 1.4 GHz / 2.4 GHz
Smart Hopping telemetry device radio is activated.
• Integrated data in a single sector at the Information Center.
Restrictions
• The MP2, X2, or MP5 cannot be connected to a host monitor (Companion mode), and cannot use
IntelliVue Instrument telemetry (IIT).
• While ECG Source is the telemetry device the ECG Out on the Intellivue Patient Monitor is not
functional.
Configuration: Use Model 2
Required Configuration on MP2, X2, or MP5
Set Network setting TAAP to (for wired or wireless)
Assign Equipment label
SRR (for wireless), (in Setup Hardware)
SRR Channel (for wireless), (in Setup Hardware)
Required Hardware on MP2, X2, or MP5
SRR interface or TAAP connector
IntelliVue Instrument telemetry (IIT) interface
Required Configuration at Information Center for MP2, X2, or MP5
Equipment label requirement
Required Configuration at Information Center for Telemetry
Equipment allowed
Equipment label requirement
Required Hardware for Telemetry
Short Range Radio Adapter - SRRA or TAAP Cable
Troubleshooting FAQ for Use Model 2
I am getting a message on the bottom of the screen that says " No companion mode in this conf"
or "No companion mode support in this telemetry configuration" and an
INOP MMS Unsupported, what can I do?
• This message will show anytime you connect an X2 or MP5 to a larger IntelliVue monitor. It is
configured specifically for use model 2. If you need this monitor to function as a companion monitor,
please refer to the instructions for use model 4.
I paired the telemetry device and the monitor at the IIC in sector setup but when I bring the telemetry device
with SRRA into the room, it stays on OBO and never changes to the bedside ECG, what can I do?
• The telemetry device searches for wireless short range radios in the monitor for 30 seconds on startup or
when the check button is pressed to preserve battery life. You can pair at the IIC and insert the batteries
when you get to the patient's room so it is searching while in range of the monitor or press the check
button when you get to the patient's room.
296
Telemetry Device Use Models
Enabled
yes
On
11-26 (service personnel only)
yes
IIT cannot be used
Monitor label
IntelliVue Telemetry only
Telemetry device label - ITS
yes