7 Telemetry Device and CL Pods Use Models
Restrictions
• When paired to a telemetry device, the monitor cannot be connected to the IntelliVue Information
Center via 1.4 GHz / 2.4 GHz Smart Hopping telemetry radio; it must be LAN connected or utilize
the WLAN (802.11) solution.
• While ECG Source is the telemetry device the ECG Out on the Intellivue Patient Monitor is not
functional.
• Pacer adjustments should not be made while paired unless the bedside monitor is the ECG source
(ECG delay is too long).
Configuration: Use Model 1
Required Configuration on Monitor
Set Network setting TAAP to
Assign Equipment label
Required Hardware on Monitor
SRR interface or TAAP connector
IntelliVue Instrument Telemetry (IIT)
Required Configuration at Information Center for Monitor
Equipment label requirement
Required Configuration at Information Center for Telemetry
Equipment allowed
Equipment label requirement
294
not used for this configuration, can be set to Enabled or Disabled
yes
not used
IIT cannot be used
Monitor label
SDN Telemetry, IntelliVue Telemetry (ITS)
Telemetry label - SDN or ITS as appropriate
Telemetry Device Use Models