Download  Print this page

Philips IntelliVue MX800 IntelliVue MP2 Configuration Manual Page 287

Patient monitor..
Hide thumbs

Advertisement

Configuring Profiles and Settings
Configuring CL Pod NBP
Main Setup -> NBP
Main Setup -> NBP -> Setup Sequence
IntelliVue CL Pod
Mode
Repeat Time
Phase A
every
Phase B
every
Phase C
every
Phase D
every
NBP
Pulse
VP Pressure
Reference
Unit
Done Tone
Start Time
Color
CL Pod NBP Configuration Implications
Mode With the setting Auto a measurement cycle is automatically started with the set Repeat Time.
Configure Mode to Manual to allow NBP measurements from the CL Pod to be made on request and
not in a sequence. Configure Mode to Sequence, to start measurement cycles as defined in the Setup
Sequence menu.
Repeat Time This setting defines the Repeat Time of the automatic measurement. Choices are
from 1 min up to 24 hours.
Phase A (B/C/D), every These settings are only visible if Mode is set to Sequence and you
select Setup Sequence in the Setup NBP menu. You can then define up to four measurement cycles
which will run consecutively. For each cycle you can set the number of measurements and the interval
between them. If you want to run less than four cycles in a sequence, set the number of measurements for
one or more cycles to Off.
Pulse This lets you enable (On) or disable (Off) display of the Pulse numeric derived from the NBP
measurement. If the NBP numeric area on the patient monitor screen is configured large enough, the
Pulse numeric will be displayed next to the NBP label in the NBP segment.
VP Pressure
cuff deflates automatically after a set time (adult/pediatric: 170 seconds, neonatal: 85 seconds) if it is not
manually deflated beforehand.
C M Factory Default Setting
x
x Auto
x
x 15 min
x
x 4 times
x
x 5 min
x
x 4 times
x
x 10 min
x
x 4 times
x
x 15 min
x
x 4 times
x
x 30 min
x
x On
x
x On
x
60 mmHg
Auscultatory (not changeable)
x
mmHg
x
Off
x
Synchronized
x
Red
This setting determines the cuff pressure used during a Veni Puncture inflation. The
6 IntelliVue Cableless Measurements