Configuring Profiles and Settings
Modifying a Setting
To change settings in an existing Profile:
1
Select the Profile (A-D) you want to change and select the Load pop-up key to activate it.
2
Make the changes to the individual measurements or CL Pod settings.
Select Profile again and select Store to save the settings to this Profile.
3
4
To reset the selected Profile to factory defaults, select Reset To Factory.
Changing the CL Pod's Default Profile
Every CL Pod has one default Profile (Profile A). This is marked with a black diamond.The CL Pod loads
the default Profile:
• when you exit Demonstration mode (but not after exiting Configuration or Service mode).
• after discharging a patient.
• after being switched off for more than 60 seconds (only if Automat. Default is set to Yes).
• when selecting New Patient or Free Device from the Patient menu.
Profile A is always the default Profile.
Configuring CL Pod SpO
Main Setup -> SpO
IntelliVue CL Pod
Mode
Repeat Time
Label
SpO
2
Pulse
Signal Quality
Average
Average in Mon.
NBP Alarm Suppr.
Color
CL Pod SpO
Mode Configure Mode to Manual to allow SpO
request and not continuously; helping to save the CL Pod's battery power. When you change to Auto, a
measurement cycle is automatically started with the set Repeat Time. When date and time are
available, the measurement is synchronized to the clock.
2
2
C M Factory Default Setting
x
x Continuous
x
x 15 min
x
x SpO
2
x
x On
x
x On
x
On
x
10 sec
x
No
x
On
x
Cyan
Configuration Implications
2
6 IntelliVue Cableless Measurements
measurements from the CL Pod to be made on
2