Configuring Special Screen Settings
Change View The screen trend presentation can be configured to Tabular, Graphical,
Horizon, or Histogram. The Tabular view can only be used with aperiodic measurements, such as
NBP, C.O., C.I., PAWP. If you configure the view of an NBP trend, for example, to Tabular, and
during monitoring the user changes the trend to a periodic measurement, such as ABP, the view
automatically switches to Graphical.
ShowHorizon Trend The horizon view is made up of 4 elements:
a horizon, drawn in white, as a reference baseline to help you visualize changes in the patient's
1
condition.
a graphical trend, displaying patient data for the set TrendTime.
2
3
a trend indicator arrow, indicating how the patient trend has developed in the last ten minutes.
a deviation bar, showing how the currently measured value deviates from the stored baseline.
4
If you set ShowHorizon Trend to Yes, all 4 elements of the horizon view are shown. If you set it to
No, the graphical trend information (2) is not displayed in the trend channel.
Configuring an Embedded Trend Window
To change the following settings, select the embedded Trend window on the Screen to open the related
pop-up key line. Note that the following settings are Screen settings and therefore do not affect the
behavior of the normal (not embedded) Vital Signs and Graph Trend windows.
Select Interval This setting defines the trend interval that will be used in the embedded trend
window when the corresponding Screen (with the embedded trend window) is opened.
Graph Trend / Vital Signs This setting defines the view (Graphical Trends or Vital Signs
table) of the embedded trend window when the corresponding Screen is opened. The view can be changed
at any time.
Select Group This setting defines the trend group that is displayed in the embedded trend window
when the corresponding Screen is opened.
The following setting applies for embedded Graph Trend windows only. To change the setting, select the
embedded Graph Trend window on the Screen, then select one of the segments on the left side of the
window to open the segment menu.
No. of Segments This setting defines the number of trend segments displayed in the embedded
Graph Trends window when the Screen is opened.
Configuring an Embedded CSA Window
To change the following settings, select the embedded CSA window on the Screen to open the related
MX800
pop-up key line. Note that the following settings are Screen settings and therefore do not affect the
MP40-90
behavior of the normal (not embedded) CSA window.
only
On/Off SEF defines whether the SEF trend line is displayed in the embedded CSA window when the
corresponding Screen is loaded.
On/Off MDF defines whether the MDF trend line is displayed in the embedded CSA window when the
corresponding Screen is loaded.
On/Off PPF defines whether the PPF trend line is displayed in the embedded CSA window when the
corresponding Screen is loaded.