Download  Print this page

Philips IntelliVue MX800 IntelliVue MP2 Configuration Manual Page 178

Patient monitor..
Hide thumbs

Advertisement

4 Configuration Settings Appendix
If you are using a barcode reader to enter the Quick Admit data, make sure that the setting made here
corresponds to the information provided by the barcode.
The settings are greyed out and not accessible in this menu if corresponding Demographic Fields are set to
Required or Hidden.
• They are automatically switched On if they are Required in the Demographic Fields settings.
• They are automatically switched Off if they are Hidden in the Demographic Fields settings.
• At least one name field or one Id field must be visible.
Last Name, First Name, Middle Name, Encounter Id, Patient Cat. Paced,
Date of Birth Configure these settings to Off, if you want them to appear in the Quick Admit
window.
Lifetime Id Configure Lifetime Id to Off, if you want to switch it off in the Quick Admit
window.
QuickAdmitDischg When a Quick Admit is initiated, the monitor compares the information
entered into the Quick Admit fields (see above) with the information that is currently stored for that field.
If the information is different, the monitor optionally discharges the previous patient, depending on how
the QuickAdmitDischg setting is configured:
– If set to Ask User, the monitor asks the user whether they want to discharge the previous patient.
– If set to Yes, the monitor automatically discharges the previous patient.
– If set to No, the monitor overwrites the data in the Quick Admit Field with the new data, but does
not discharge the patient.
• At least one name field or one Id field must be visible.
N O TE
• If the monitor is connected to an Information Center, the monitor enters "---" into the Last
Name, Lifetime Id, and Encounter Id fields to enable admission at the Information Center.
• Check the Quick Admit settings after upgrading or downgrading.
Configuring Remote Display Settings
MX800
Global Setting: Main Setup -> Global Settings -> Remote Display
MP60-90
MP2/X2
Factory Defaults
MP5
Item Name
only
Access Rights
Password
Standby Image
Remote Display Configuration Implications
These settings manage aspects of the connection of an IntelliVue monitor to the XDS Remote Display.
178
Oper.
MX800,
Mode
MP60 - MP90
C M
MP2/X2
MP5
x
Operating
x
0000
x
Boot Image
Global Settings

Advertisement