4 Configuration Settings Appendix
TeleUnassign This setting is only available in monitors that have a short range radio interface
installed. It determines whether the monitor will automatically break the assignment to a telemetry device
connected via a short range radio connection under certain circumstances.
• If configured to 1 min, the monitor will automatically unassign a telemetry device from the monitor
when the monitor
– is switched off for longer than 1 minute, or
– is placed in Standby mode for longer than 1 minute while having no connection to the IntelliVue
Information Center and the telemetry device.
• If configured to Never, the monitor will not automatically break the assignment.
For general information about the configuration of the IntelliVue Information Center and Telemetry
Devices refer to their Configuration Guides.
For specific information about the configuration for different use models refer to:
Telemetry Use Models (page 297)
Cableless (CL Pod) Measurements Use Models (page 307).
Configuring Auto Discharge Settings
Global Setting: Main Setup -> Global Settings -> Auto Discharge
Factory Defaults
Item Name
Power Off
Standby
No Basic Vitals
Auto Discharge Configuration Implications
Power Off can be configured to a specific time (1min, 10min, 30min, 1h, 3h, or 8h) or to
Never. If the monitor is switched on, after being switched off for longer than the specified time, it will
automatically discharge the current patient and begin monitoring a new patient
Standby can be configured to a specific time (1min, 10min, 30min, 1h, 3h, or 8h) or to Never. If
monitoring is resumed, after the monitor was in Standby for longer than the specified time, it will
automatically discharge the current patient and begin monitoring a new patient
No Basic Vitals can be configured to a specific time (1min, 10min, 30min, 1h, 3h, or 8h) or
to Never. If no basic vitals (HR, RR, Pulse, SpO
monitor will automatically discharge the current patient and begin monitoring a new patient
174