Global Settings
• the patient in the MMS and the patient in the monitor are different, and Pat.Sel.Default is
configured to Continue MMS.
• the patient in the MMS and the patient in the monitor are different, and Pat.Sel.Default is
configured to Ask User, and the clinician resolves the patient identification mismatch by selecting
Same Patient or Cont MMS.
Note: if Pat.Sel.Default is set to Cont Monitor, MMS PW Sync is automatically set to No
and cannot be changed.
It is not possible to transfer SSC Sepsis Protocol data from a monitor with release F.0 software to another
NO TE
monitor with release G.0 software (or higher) and vice versa.
Tele Discharge This setting is only available in monitors with a telemetry device (TAAP) connector
or a short range radio interface. It determines the automatic discharge behavior of the monitor only when
it is used in a mode where it:
– has no connection to a host monitor (companion mode),
– has no connection to an Information Center, and
– a telemetry transceiver is directly connected via cable or short range radio link to the monitor.
The available configuration choices are:
• OnDisconnect The monitor automatically discharges the patient from the monitor when the
telemetry transceiver is disconnected or unassigned from the monitor.
Be aware that if you then measure NBP or Pred. Temp with the monitor before you connect a new
transceiver, only the measurement readings with the latest timestamp will be uploaded to the
Information Center when you connect the new transceiver.
• OnDevChange The monitor automatically discharges the patient from the monitor when a different
transceiver is connected or assigned to the monitor. The transceivers are distinguished by their
telemetry labels. If you connect the transceiver to the monitor via a short range radio link, assigning a
different transceiver automatically unassigns the previous transceiver.
With Tele Discharge configured to OnDevChange, be aware of the following: if you want to
measure NBP or Pred. Temp for a patient and want to upload this data via the transceiver to the
Information Center, always connect the transceiver to the monitor before you make the measurements,
otherwise the measurement data will be erased by the discharge when you connect the new transceiver.
• Off This setting is not recommended if you want to use the monitor for spot-checking. With Tele
Discharge set to Off, the monitor does not perform an automatic discharge.
If you disconnect a transceiver and then measure NBP or Pred. Temp with the monitor, the latest
measurement readings will be uploaded to the Information Center when you connect a new transceiver.
Height Unit/ Weight Unit define the unit used when entering the height / weight of the patient.
Choices are in or cm for height and lb or kg for weight. Be aware that these settings can be overwritten
by the Region settings applied during an Upgrade or Cloning procedure with the IntelliVue Support Tool.
Automat. Default
– If Automat. Default is set to Yes, and the monitor is switched off for more than one minute,
the default Profile is reloaded in the monitor. Any unstored changes made to the Settings Blocks and
Profiles are lost.