Philips IntelliVue MX800 IntelliVue MP2 Configuration Manual Page 170

Patient monitor..
Hide thumbs

Advertisement

4 Configuration Settings Appendix
Asystole Detect. Set Asystole Detect. to Enhanced to improve alarming on asystole
under certain conditions. In enhanced mode, an asystole alarm will be suppressed for up to nine seconds (=
Asystole Threshold + 5 sec.) if a valid beat-to-beat Pulse is detected from an active pulsatile invasive
Pressure measurement.
Pat.Sel.Default This setting defines how the monitor behaves when there is a patient
identification mismatch between the MMS and the monitor. If you set this to Cont Monitor or
Continue MMS, the monitor resolves the mismatch automatically. To require user confirmation before
the mismatch is resolved, set this to Ask User. For more detail, please refer to the monitor's Instructions
for Use.
AskUser Reminder This setting lets you specify a time after which the clinician will be reminded of
an unresolved patient identification mismatch. If the user ignores a patient identification mismatch by
closing the Patient Selection window, the monitor will automatically pop up the Patient Selection window
on the Main Screen after the configured time (5min, 10min, 15min, or 30min). To disable this
behavior, set AskUser Reminder to Off.
MMS Sett.Upload If set to Yes, the active settings from the MMS measurements will be uploaded
to the monitor when you connect an MMS to a monitor and one of the following conditions applies:
• the patient in the MMS and the patient in the monitor are the same, i.e. no patient identification
mismatch occurs.
• the patient in the MMS and the patient in the monitor are different, and Pat.Sel.Default is
configured to Continue MMS.
• the patient in the MMS and the patient in the monitor are different, and Pat.Sel.Default is
configured to Ask User, and the clinician resolves the patient identification mismatch by selecting
Same Patient or Cont MMS.
Note: if Pat.Sel.Default is set to Cont Monitor, MMS Sett.Upload is automatically set
to No and cannot be changed.
MMS Trend Upload If set to Yes, the trend data from the MMS measurements will be uploaded to
the monitor when you connect an MMS to a monitor and one of the following conditions applies:
• the patient in the MMS and the patient in the monitor are the same, i.e. no patient identification
mismatch occurs.
• the patient in the MMS and the patient in the monitor are different, and Pat.Sel.Default is
configured to Continue MMS.
• the patient in the MMS and the patient in the monitor are different, and Pat.Sel.Default is
configured to Ask User, and the clinician resolves the patient identification mismatch by selecting
Same Patient or Cont MMS.
Note: iff Pat.Sel.Default is set to Cont Monitor, MMS Trend Upload is automatically set
to No and cannot be changed.
MMS PW Sync If set to Yes, the ProtocolWatch information stored in the MMS (PW data, PW state,
PW logs and active protocol) will be uploaded to the monitor when you connect an MMS to a monitor
and one of the following conditions applies:
• the patient in the MMS and the patient in the monitor are the same, i.e. no patient identification
mismatch occurs.