Global Settings
Factory Defaults
Item Name
ConfirmAlarmsOff
Power Loss Sound
Label Set
LAN Data Export
4
TeleUnassign
2.
TAAP
Remote Display
Setup Internal PC
1.For new monitors shipped from the factory, the defaults for these settings are set for the country to which the monitor is
shipped, see "Configuring Country-Specific Settings" on page 186.
2.Setting only available for monitors that have a telemetry device (TAAP) connector or a short range radio interface installed.
3.MP5SC only
4.Setting only available for monitors that have a short range radio interface installed.
General Global Settings Configuration Implications
Default Profile To set the default Profile, select Profiles in the Monitor Info Line, select Profile in
the Profiles window, then select the Profile you want to set as default Profile from the pop-up list. Select
the Set Default softkey. This change takes effect immediately and you do not need to save it. Use the
table in the section "Profile Settings" on page 32 to document the default Profile.
Altitude Some measurements use the configured altitude setting to derive a typical ambient pressure
which is used in the calculation of partial pressure values. To ensure correct measurement values, the
altitude setting must be correctly set.
Line Frequency Use the Line Frequency setting to configure the correct line frequency for the
AC Power, either 50 Hz or 60 Hz. If the Line Frequency is incorrectly set, this may affect the ECG signal
quality.
QRS Type Select QRS Tone or QRS Tick. If Tone Modulation is set to Yes, the QRS Type
automatically switches to QRS Tone. For both types, the frequency and rhythm information is derived
from either the ECG or Pulse, depending on which is currently selected as the alarm source.
ECG Cable Color This setting determines the labels the monitor uses when it refers to individual
ECG electrodes, such as in a LEAD Off INOP message that will be issued when an individual ECG
electrode has fallen off. If ECG Cable Color is set to AAMI, the monitor uses the labels RA, LA, LL,
RL, V, and V1 through V6. If set to IEC, it uses RA, LA, LL, RL, and C1 through C6.
Oper.
MX800,
MX800,
Mode
MP20 - MP90
MP20 -
(H10/20/40)
MP90(H30)
C M
x
No
x
not applicable, setting is not
available for MX800, MP2-90
x
Restricted
x
All
x
not applicable, setting is not
available for MX800, MP20-90
x
not applicable, setting is not
available for MX800, MP20-90
x
See "Configuring Remote Display Settings" on page 178
x
MX800 only; See "Configuring Remote Display Settings" on page 178
4 Configuration Settings Appendix
MP5
MP5
MP5T,
(H30)
(H10/20/40)
MP5SC,
MP5#P05
Enabled
1 min
Enabled
MP2 / X2
not
applicable
MP2:
Enabled
X2: Disabled
169