4 Configuration Settings Appendix
When you clone a configuration to a monitor, the associated Guardian Early Warning Scoring protocol is
WARNING
automatically cloned with the configuration. Therefore, before cloning a configuration to a monitor, you
must always check
• whether there is a Guardian Early Warning Scoring protocol in the configuration
• whether this configuration is correct for the hospital unit you are working in
• that you have a Guardian Early Warning Scoring protocol offline configuration report on file, signed by
a hospital representative.
For detailed information on configuring Guardian Early Warning Scoring protocols, see the chapter
"Using the EWS Configuration Editor" in the Support Tool Instructions for Use.
Configuring Recordings
Unique Monitor Setting: Main Setup -> Recordings
Factory Defaults
Item Name
General
Central Config
ECG Gain
Delayed Recording
Recording Name
Recorder
Channel 1
Channel 2
Channel 3
1
Channel 4
Overlap
Speed
Delay Time
Run Time
Alarm Recording
Recorder
Channel 1
Channel 2
Channel 3
Overlap
Speed
Delay Time
154
Oper.
MX800,
Mode
MP20 - MP90
C M
MP5 (H30)
x
x No
x
x 10 mm/mV
x
Recordng
x
x Local
x
x Primary Lead
x
x ABP
x
x Blank
x
x Blank
x
x Off
x
x 25 mm/s
x
15 sec
x
20 sec
x
x Local
x
x Primary Lead
x
x Alarm Par
x
x Blank
x
x Off
x
x 25 mm/s
x
15 sec
MP20 (M20/M21)
MP5 (H10/20/40)
MP5T, MP5SC
Unique Monitor Settings
MP2/X2
Central 2-Ch
Central 2-Ch