Monitor Settings
4 Configuration Settings Appendix
IIT This setting is not a monitor setting, it is a global setting. It is available with IntelliVue X2, MP2,
MP5, MP20/30 and MP40/50 monitors only. The Instrument Telemetry network interface can be
disabled to suppress network related technical INOPs if the device is operated in an infrastructure without
telemetry.
RF Access Code This setting is not a monitor setting, it is a global setting. It is available with
IntelliVue X2, MP2, MP5, MP20/30 and MP40/50 monitors only. It is needed to establish
communication of the IntelliVue Instrument Telemetry transceiver with the access points of the wireless
network. The IntelliVue Instrument Telemetry transceiver only communicates with an access point that is
configured to the same RF Access Code.