Monitor Settings
Auto Fill Waves If Auto Fill Waves is set to Yes, the monitor automatically assigns
available waves that would otherwise not be displayed, to wave channels for which the preconfigured waves
are currently not available. Set Auto Fill Waves to No, if you do not want waves other than the ones
preconfigured to automatically appear on the Screen.
An example: assume you are viewing a 4 wave Screen with the following wave channel configuration (top
to bottom): ECG, Any SpO
currently not measuring CO
Invasive Pressure to the 4th wave channel. With Auto Fill Waves set to No, the 4th wave channel
would be empty.
Show ST In Wave If Show ST in Wave is set to Yes, the current ST value will be shown next
to each ECG wave.
Hide Alarm Info This setting only applies to monitors with two or three main displays (excluding
the XDS Remote Display). It is only available for Display 2 and Display 3. If you set it to Yes,
for example for Display 3, this display will no longer show any alarm related information, such as
alarm messages, alarm limits next to the numerics, and Alarm off symbols. Flashing numerics will also be
deactivated on this display.
Note that this setting has no effect on the XDS Remote Display. The monitor will behave as if the setting
was set to No.
Hide Keys This setting only applies to monitors with two or three main displays (excluding the XDS
Remote Display). It is only available for Display 2 and Display 3. If you set it to Yes, for example
for Display 3, the Operating Area, including permanent keys, SmartKeys, and the Measurement
Selection key will be deactivated and invisible on this display.
Note that this setting has no effect on the XDS Remote Display. The monitor will behave as if the setting
was set to No.
NBP Time If NBP Time is set to Meas Time, the time shown beside the NBP numeric will show the
timestamp of the most recent NBP measurement. If set to Next Meas, and NBP mode is set to Auto
or Sequence, the time until the next automatic measurement is shown, along with a graphic
representation of the remaining time.
Wave Style This setting lets you change the presentation of certain respiratory waveforms (CO
AWF, AWP, and AWV). If you set this to Filled, the area underneath the waves is filled with color. If
you set it to Line, the waves are displayed in the standard way, as continuous lines.
Wave Line Style This setting lets you configure the thickness of all waves and HiRes Trends on
display 1 (2,3). For better visibility over a distance you might want to use Medium or Thick. The
choices are: Thin, Medium, Thick, Extra Thick.
Display-Related Configuration Implications for Monitors which are
connected to an XDS Remote Display
• In an MX800, MP2/X2, MP5, MP60/70, MP80, MP90 (with single CPU), the XDS Remote Display
will be the second main display. The XDS Remote Display will use the settings made for Display 2.
Note that the User Interface settings Hide Alarm Info and Hide Keys made for Display 2
will have no effect on the XDS Remote Display. The monitor will behave as if the settings are set to No.
, Any Press, CO
. With Auto Fill Waves set to Yes, if you were
2
2
but a 2nd Invasive Pressure, the monitor would automatically assign the 2nd
2
4 Configuration Settings Appendix
,
2
131