4 Configuration Settings Appendix
Monitor Settings
Meas. Deactivate This setting defines the monitor's behavior when you connect a measurement
device (MMS, X2, or plug-in module) to the monitor, and
• a measurement label of the newly connected device duplicates the label of an already connected
measurement device, and
• the already connected device is switched off.
If Meas. Deactivate is set to Manual, the monitor deactivates the newly connected measurement,
due to the label conflict.
If Meas. Deactivate is set to Auto, the monitor deactivates the previously connected measurement
device and activates the newly connected device.
Timer Volume determines the volume of the notification tone for all timers.
Companion - Operating This setting is only available on host monitors that can support the X2
or MP5 as a multi-measurement module (companion mode). It determines whether the companion
monitor (X2 or MP5) can still be operated, when it is connected to a host monitor. If Enabled,
functions you can operate on an X2 or MP5 are restricted to measurements originating in that device. If
Disabled, no operation is possible.
Companion - Display This setting is only available on host monitors that can support the X2 or
MP5 as a multi-measurement module (companion mode). It determines which Screen is shown on the
companion monitor (X2 or MP5), when it is connected to a host monitor. Choices are:
• Curr. Screen: shows the Screen that was active before connecting to the host.
• Blank Screen: shows a special companion mode standby Screen.
Brightness Defines the default brightness for monitoring. This setting is not applicable for an MP80/
90 monitor (external display). For Brightness, the choice Optimum is equivalent to 8.
Lets you choose a brightness setting for when the monitor is in Standby.
Standby Brightness
This setting is not applicable for an MX800, MP80/90 monitor (external display). For Standby
Brightness, the choice Optimum is equivalent to 1.
Transport Brightness This setting is only available on the MP2 - MP50. It defines the display
brightness when the monitor is running on battery power. For Transport Brightness, the choice
Optimum is equivalent to 4.
User Interface - Display Settings Configuration Implications
Use this to change the line spacing for all menus. Wide provides more room
Menu LineSpacing
between menu entries and is the recommended setting for touchscreen use. Narrow lets you see more
menu entries on one menu page and is recommended when using mouse or SpeedPoint/Navigation Point
as the primary input device.
Display Units If Display Units is set to Yes, the measurement units are displayed next to the
measurement numerics, provided that there is enough space.
Alarm Limits If Alarm Limits is set to Yes, the alarm limits are displayed next to the
measurement numerics, provided that there is enough space.
130