Download  Print this page

Philips IntelliVue MX800 IntelliVue MP2 Configuration Manual Page 123

Patient monitor..
Hide thumbs

Advertisement

Monitor Settings
Auto Reports Configuration Implications
Auto Reports print automatically when a specified trigger occurs. There are two types of Auto Reports:
• Scheduled reports can be set up to print at predefined intervals, starting at a predefined time of
day. The start time you set applies for every following day. For example, if you set a start time of 07:00
and a repeat time of six hours, the first report will print at 07:00 every day, the next at 13:00 and so on.
• End Case Reports print when a patient is discharged using the monitor's End Case function.
An Auto Report can include both a Scheduled Report and an End Case report.
Setting up an End Case Report
Select Main Setup -> Reports -> Auto Reports
1
Select one of the four available Auto Report "slots" (A, B, C, or D)
2
3
Select Report, then select the type of report you want to be printed when a patient is discharged
using the End Case function, for example "VitalsReport".
Set End Case Report to On.
4
5
Set Scheduled Rep. to Off, if do not want the monitor to trigger a Vital Signs Report at
predefined intervals.
If needed, set up the VitalsReport or whatever type of report you used for the EndCase Report.
6
7
If you want more than one report to be printed when using End Case on a patient, repeat steps 2
through 6, selecting a different Auto Report (A, B, C, or D) and allocating it a different report
type.
Configuring ECG Reports
ECG Reports Configuration
Not
MP5T
Monitor Setting: Main Setup -> Reports -> ECG Reports
MP5SC
Factory Defaults
Item Name
ReportLead Layout
ECG Gain
Speed
Annotation
ECG Reports Configuration Implications
Report Lead Layout The layout of the report can be configured to either International or
Cabrera.
ECG Gain This lets you set a defined ECG Gain to determine how ECG waves will appear on the ECG
report printouts.
Oper.
MX800,
Mode
MP2 - MP90
C M
X2
x
Internat
x
x 10 mm/mV
x
x 25 mm/sec
x
x Off
4 Configuration Settings Appendix