Download  Print this page

Philips IntelliVue MX800 IntelliVue MP2 Configuration Manual Page 120

Patient monitor..
Hide thumbs

Advertisement

4 Configuration Settings Appendix
– Alarm Limits: a list of all currently set alarm limits.
– Battery Stat: Battery Status report.
Report Size Lets you choose the paper size to be used for the report. The selection of sizes listed
depends on the report type selected: Unspecified, Universal, A4, Letter, LrgUniversal,
A3, or Ledger.
• If set to Unspecified, the report prints on the first available printer in the list of printers which is
enabled and which has a paper size appropriate for the chosen template.
• If set to Universal, the report prints on the first available printer in the list of printers which is
enabled and which offers the paper size A4 or US letter.
• If set to LrgUniversal, the report prints on the first available printer in the list of printers which is
enabled and which offers the paper size A3 or ledger.
Orientation lets you set the orientation of the report printout. The selection of orientations listed
depends on the report type and report size selected:
• Unspecified: the report uses the default orientation for the chosen report type.
• Landscape: the report uses landscape orientation.
• Portrait: the report uses portrait orientation.
Target Device Lets you choose which printer the print job will be sent to:
• If set to Unspecified, the report prints on the first printer in the list of printers which is enabled
and which has a paper size appropriate for the chosen template. If no printer is available when the report
is triggered, the report is sent to the print database from where it will be automatically printed, when an
appropriate printer becomes available.
• If set to Local 1 / 2, the report will be sent to the corresponding local printer. If the printer is not
available when the report is triggered, the report is not printed, and the information is lost.
• If set to Remote 1 / 2 /3, the report will be sent to the corresponding remote printer. If the
printer is not available when the report is triggered, the report is not printed, and the information is
lost.
• If set to Database, the report will be sent to the print database (provided this is enabled, see "Print
Database" on page 209). The print database is a special section of the monitor database which acts as a
buffer for print jobs. Reports stored in the print database will not be cleared by a discharge or by a
power cycle. Reports stored in the print database will be handled depending on how the setting
Auto Prnt Dbs is set, see "Configuring Other Report Settings".