4 Configuration Settings Appendix
ProtocolWatch Application Configuration Implications
ProtocolWatch (PW) is a clinical decision support tool. It allows you to run a clinical protocol which can
monitor developments in the patient's condition. ProtocolWatch notifies you when certain conditions or
combinations of conditions occur and it documents developments in a log which can be printed. For
detailed information on how to use this application, see the monitor's Instructions for Use.
CVP Threshold This setting lets you define the maximum CVP value the monitor will automatically
enter into a protocol when CVP is continuously measured. Use this setting to prevent false high CVP
values from being entered into a protocol when the CVP is continuously measured.
CVP Thresh. Unit This setting lets you change the unit used for the CVP Threshold.
Protocol This setting lets you preselect the protocol you want to run. Selecting the Protocol also
determines which protocol-specific configuration settings are displayed (see below) and can be configured.
At this point, the following protocols are available:
• SSC Sepsis Protocol
• Guardian Early Warning Scoring Protocols:
– SpotCheck
– EWS SPS
If you select Off, no protocol will be preselected. Note that Protocol can always be changed by the
clinician during monitoring.
Configuring SpotCheck Monitoring
MP5SC,
Main Setup -> SpotCheck-> Automatic VS
MP5
only
Factory Defaults
Item Name
Interval
Interval Time
NBP
SpO2
Patient Alarm
1.Profile SpotCheck for English language software only.
These settings are profile dependent monitor settings. They only take effect if currently either
SpotCheck (see "Configuring Guardian Early Warning Scoring" on page 153) or EWS SPS protocol
(see "Configuring Recordings" on page 154) is executed.
110
Oper.
MP5SC, MP5#P05
Mode
C M
Adult,
EWS Scoring,
SpotCheck
Pedi,
Neo
x
x No
x
x 1 hrs
x
x No
x
x No
x
x No
Profile
Frequent
RRT
1
Vitals
Yes
2 min
Yes
No
Yes
No
Yes
No
Monitor Settings
Resus

Advertisement