Download  Print this page

Philips IntelliVue MX800 IntelliVue MP2 Configuration Manual Page 109

Patient monitor..
Hide thumbs

Advertisement

Monitor Settings
Configuring ST Map
Main Setup -> ST Map
Factory Defaults
Item Name
Interval
Scale
ST Map Settings Configuration Implications
Interval From the ST Map window, use the pop-up key Select Interval to configure the time
interval between the trended ST map samples when using ST Map in Trend View. The interval ranges
from 12 seconds to 30 minutes.
Scale In the ST Map window, switch to Trend View and use the Size Up / Size Down keys to
define the scale that the monitor uses in the ST Map application. Available choices are 1mm, 2mm, 3mm,
5mm, 10mm, and 15mm.
Configuring ProtocolWatch
MX800
Main Setup -> ProtocolWatch
MP5-90
only
Factory Defaults
Item Name
CVP Threshold
CVP Thresh. Unit
Protocol
1.Profile SpotCheck for English language software only.
Oper.
MX800,
Mode
MP2 - MP90
C M
MP5T, MP5SC
X2
x
x 5 min
x
x 2 mm
Oper.
MX800,
Mode
MP5 - MP90
MP5SC
C M
C M
x
100 mmHg
x
mmHg
x
x Off
4 Configuration Settings Appendix
MP5#P05
MP5SC
Profile
SpotCheck EWS SPS
109

Advertisement

Table of Contents

Comments to this Manuals

Symbols: 0