Philips IntelliVue MX800 IntelliVue MP2 Configuration Manual Page 108

Patient monitor..
Hide thumbs

Advertisement

4 Configuration Settings Appendix
Configuring Vital Signs Recording
MX800
Monitor Setting: Main Setup -> Recordings -> Vital Signs
MP5-90
only
Factory Defaults
Item Name
End Case Record.
Trend Group
Period
Column
Interval
Show Unit
Vital Signs Recording Configuration Implications
End Case Record. Set this to On if you want the Vital Signs recording be triggered when a patient
is discharged using the End Case function.
Trend Group. Use this setting to define the default trend group that is recorded each time a Vital
Signs recording is triggered.
Period. Use this setting to define the period of time for which trend data should be printed on the
recording. Available choices are: 1, 2, 3, 6, 12, 24, 48, 72, and 96 hours.
Column. Use this setting to define in which way the Vital Signs recording is displayed on the recorder
paper.
• To display the trended values with the time stamps for the set Interval time on top of the table
select Interval.Each measurement of the set Trend Group is displayed in one row.
• To display it with the time stamps left of the printed table select NBP. Each measurement of the set
Trend Group is displayed in one column.
Interval. Use this setting to determine the default time interval (resolution of the trend data) that is
recorded each time a Vital Signs recording is triggered. Available choices are: 12 sec, 1 min, 5 min,
10 min, 15 min, 30 min, 1 hours, 2 hours, 3 hours.
Show Units. If Show Units is configured to On, the measurements in the Vital Signs recording
are printed together with their units.
108
Oper.
MX800,
Mode
MP20 - MP90
C M
MP5 (H30)
x
x Off
x
x All
x
x 6 hours
x
x Interval
x
x 30 min
x
Off
Monitor Settings
MP20 (M20/M21)
MP5 (H10/20/40)
MP5T, MP5SC
On
3 hours
10 min