Monitor Settings
Monitor Setting: Main Setup -> Trends -> Horizon Settings
Factory Defaults
Item Name
Label
Horizon High
Horizon Low
Scale Delta
Unit
Horizon Trend Configuration Implications
Unit Choose the measurement unit that you will use to define the values for Horizon High,
Horizon Low and Scale Delta.
Horizon High/Low Set Horizon High to select the upper horizon value. Set Horizon Low to
select the lower horizon value. If the high and low horizon values are the same, the horizon is a baseline, if
the values are different, the horizon is a range.
Scale Delta defines the distance between the horizon and the upper and lower scale limits. Note that
the upper and lower scale limits may also be restricted by the measurement ranges defined for a label.
Make sure the values for Horizon High/Low and Scale Delta match the Unit you enter.
Be aware that the master unit of a label is defined in its measurement setup menu. For example, for the
label Temp this would be the Setup Temp menu (Main Setup -> Measurements -> Temp).
If for a label, the unit defined in the Horizon Settings > Setup Measurement menu differs from the master
unit defined in the related measurement setup menu, the values defined for Horizon High/Low and
Scale Delta will be automatically converted when you close the Horizon Settings -> Setup
Measurement menu to match the unit defined in the measurement setup menu.
Example: In the Horizon Settings window, you add the label "Temp". You set the Unit to
Horizon High and Horizon Low to 35.0
o
menu to
F, the Horizon High/Low will be automatically converted to the equivalent of 35
Fahrenheit which would be 95.0
Configuring Global Trend Style
Monitor Setting: Main Setup -> Trends
Factory Defaults
Item Name
Global Style
Oper.
MX800,
Mode
MP2 - MP90
C M
MP5T, MP5SC
X2
x
<none>
x
<none>
x
<none>
x
<none>
x
<none>
o
F.
Oper.
MX800,
Mode
MP2 - MP90
C M
MP5T, MP5SC
X2
x
x Band
4 Configuration Settings Appendix
o
C. If you then change the Unit in the Setup Temp
o
C, and
o
C in
105