4 Configuration Settings Appendix
Configuring Auto Alarm Limits
Monitor Setting: Main Setup -> Alarms -> Setup AutoLimits
Factory Defaults
Item Name
All
HR
ST
SpO
2
Pulse
Press
NBP
CO
2
awRR
O
2
RR
tcGas
Temp
N
2
N
O
2
Agent
Alarm AutoLimits Settings Configuration Implications
All Set this to Disabled if you do not want the user to be able to apply AutoLimits to all enabled
measurements at once by using the All Lim. Narrow or All Lim. Wide pop-up keys in the
Alarm Limits window.
<Measurement> If you set a <Measurement> to Disabled, AutoLimits cannot be applied to
this measurement.
102
Oper.
MX800,
Mode
MP2 - MP90
C M
MP5T, MP5SC
X2
x
Enabled
x
Enabled
x
Enabled
x
Enabled
x
Enabled
x
Enabled
x
Enabled
x
Enabled
x
Enabled
x
Enabled
x
Enabled
x
Enabled
x
Enabled
x
Enabled
x
Enabled
x
Enabled