ResMed HumidAire 2i – Disinfection Guide This Disinfection Guide is intended for multipatient use of the HumidAire 2i in a sleep lab, clinic or hospital. If you use the water chamber as a single user in the home, refer to the User Guide for cleaning instructions.
Page 3
1 Cleaning: ResMed has tested the detergent Neodisher MediZym™ (diluted at 0.5%) using warm water (approximately 113°F or 45°C) according to the manufacturer’s instructions. 2 Disinfection: Due to specific regional requirements, ResMed water chambers have been tested for disinfection for 20 cycles using hot water (approximately 199°F or 93°C) for 10 minutes.
Page 4
Storage Storage temperature: -4°F to 140°F (-20°C to 60°C). 1 Cleaning: ResMed has tested the detergent Alconox (diluted at 1%) using hot water (approximately 140°F or 60°C) or warm water (approximately 120°F or 50°C) according to the manufacturer’s instructions. 2 Disinfection: ResMed water chambers have been validated for 20 cycles using solutions of ortho-phthalaldehyde (0.55% CIDEX OPA for 12 minutes) according to the manufacturer’s instructions.
Page 5
Disassembling and Reassembling the HumidAire 2i WARNING Beware of electrocution. Do not immerse the device or power cord in water. Always unplug the device before cleaning and be sure that it is dry before plugging back in. Turn the main power switch at the back of the device to OFF, and unplug the power cord from the power outlet.
Page 6
1. Push down on the tub plate from inside the chamber to remove it from the chamber. 2. Remove the O-ring from the tub plate. 3. Turn the water chamber base upside down. Firmly pull the latch up and away from the chamber until separated from the base.
Page 7
Clinical Use Only 2. Take the assembled catches and fit them to the water chamber base. Turn the water chamber base upside down. Insert the latch’s feet into the locking tabs. Push firmly and click into place. Refit the O-ring in the channel of the tub plate. CAUTION Ensure that you replace the O-ring if it is nicked or punctured.
ResMed Ltd (Manufacturer) 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia, ResMed Corp (US Designated Agent) 14040 Danielson Street Poway CA 92064-6857 USA, ResMed (UK) Ltd (EU Authorized Representative) 96 Milton Park Abingdon Oxfordshire OX14 4RY UK, ResMed Offices Australia, Austria, Brazil, China, Finland, France, Germany, Hong Kong, India, Japan, Malaysia, Netherlands, New Zealand, Norway, Singapore, Spain, Sweden, Switzerland, UK, USA (see www.resmed.com for contact details).