Available Service Port Connection - Emerson Micro Motion 5700 Installation Manual

Hide thumbs Also See for Micro Motion 5700:
Table of Contents

Advertisement

December 2014
8

Available Service Port connection

WARNING!
If the transmitter is in a hazardous area, do not remove the housing cover while power is being
supplied to the unit. Removing the housing cover while power is supplied to the unit could cause an
explosion. To access the service port in a hazardous environment, be sure to remove power from
the transmitter and wait 5 minutes before removing the housing cover.
You can interface with the transmitter through the service port connection,
located under the Service Port warning flap at the wiring access points. To
interface with the service port, you can use commonly available USB
hardware, such as a USB drive or USB cable. Use the service port connection
to download or upload data from/to the transmitter.
Quick Installation Guide
Installation Manual
53

Hide quick links: