Safety Certified Identification - Emerson Rosemount 5408 Reference Manual

Level transmitters non-contacting radar
Hide thumbs Also See for Rosemount 5408:
Table of Contents

Advertisement

Safety Instrumented Systems (4-20 mA only)
Figure 8-1:
A
LT
A
LT
LT
A
LT
LT
LT
A.
B.
C.
Application examples
8.4

Safety certified identification

All Rosemount 5408:SIS Level Transmitters must be identified as safety certified before
installing into SIS systems.
1.
2.
134
SIF Configuration Examples
B
B
B
Rosemount 5408:SIS Level Transmitter (sensor)
Logic-solver
Actuator
Level range monitoring
Dry-run prevention
Overfill prevention
Verify the transmitter model code starts with "5408F".
Verify the software (SW) is 1.A3 or later.
C
Single use 1oo1 (1-out-of-1) for SIL2
(SIL 2@ HFT=0)
C
Redundant use 1oo2 for SIL3
(SIL3@ HFT=1)
C
Redundant use 2oo3 for SIL3
(SIL3@ HFT=1)
Reference Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rosemount 5408:sis

Table of Contents