3.4.4: Validity
Validity of the measured value in %
This parameter provides a percentage measure of the certainty that the displayed
measured value corresponds to the real level and does not represent a multiple echo or a
fixed target. Its display can be evaluated as follows:
•
x > 70:
•
70 > x > 50:
•
50 > x > 20:
•
x < 20:
3.4.5: Sensor Temp
Sensor temperature
4. Device Data
4.1: Units
4.1.1: Length Unit = Parameter 1.2
4.1.2: Volume unit (F = m
Value
4.1.3: Mass Unit (for qualified personnel only)
This parameter is for Factory Authorized Personnel Only.
7ML19985FH04
very good
good
uncertain
no plausible measured value
3
)
bbl
3
yd
3
ft
3
in
bush
bbl (liq)
l
3
m
hL
Gal
ImpGal
SITRANS LR 400 – INSTRUCTION MANUAL