Process the batch

1 Initiate batch processing in either of the following ways:
Analysis is performed on the samples you added to the batch in the
previous topic, using parameters set in the Data Analysis method, as
described in
analysis"
2 View/check the analysis results as follows:
3 Save results using Save Analysis File As or Save from the File menu.
4 Select Generate from the Report menu, select report options, and click OK.
40
Click the Process Batch toolbar button
Select Process Batch from the Process menu.
"Create the Data Analysis (DA) method for SemiQuant data
on page 38.
a Batch Table pane- - Displays the concentration and count for each
element in the samples.
b Spectrum pane- - Displays mass spectra. You can zoom in to identify
elements by right- clicking and dragging the mouse cursor around the
desired mass number.
c Semiquant Factor pane- - Displays the SemiQuant factor graph for each
sample. Place the cursor near a point on a graph to display element
names and SemiQuant factor correction details. Use the arrow keys to
move between samples. If necessary, the factors can be changed as
described in
"Change the SemiQuant Factors"
If the SemiQuant Factor pane is not currently displayed, select
Calibration Curve from the View menu.
d ISTD Stability Graph pane- - Displays the percent recovery of each ISTD
element in the samples. See online help for more information.
G7200-90210 (Rev. A) MassHunter Workstation Quick Start Guide
, or