Reference Manual
00809-0100-4408, Rev AA
3.4.3

Threaded version

Antenna diameter (D) < Thread diameter (d)
Flanged tank connection
1.
2.
3.
4.
Mechanical Installation
PRELIMINARY
Place a suitable gasket on the tank flange.
Place the customer supplied flange over the gasket.
Tighten the bolts and nuts with sufficient torque for the flange and gasket choice.
Apply anti-seize paste or PTFE tape on threads according to your site procedures.
Gasket may be used as a sealant for adapters with 1½ or 2-in. BSPP (G) threads.
Mechanical Installation
November 2016
25